Introduction This Directions for use manual is a comprehensive guide designed to help you understand the capabilities and operation of the ProBP 3400 non-invasive blood pressure device. Page 1 Welch Allyn Connex® ProBP™ 3400 digital blood pressure device Directions for use.; Page 2 Welch Allyn. Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual. View and Download Welch Allyn Connex ProBP 3400 outline online. Digital Blood Pressure Device. Connex ProBP 3400 Blood Pressure Monitor pdf manual download. Welch Allyn Connex® ProBP™ 3400 digital blood pressure device. Intended use published in this manual. Welch Allyn, SureBP, Connex, and FlexiPort are registered. Quick Overview. Small enough to fit in your hand, with the same accuracy as Welch Allyn industry-leading vital signs devices, the ProBP 3400 digital blood pressure device is a smart choice for clinicians in just about every medical setting.
Advertising
1. Connect a USB cable to the service test box; connect the opposite end to a PC.
2. Connect a mini B USB cable to the ProBP 3400 and the opposite end to the service test box. 2004 samsung 32 inch tv user manual pdf.
2. Connect a mini B USB cable to the ProBP 3400 and the opposite end to the service test box. 2004 samsung 32 inch tv user manual pdf.
3. Connect the blood pressure hose to the NIBP module and the test volume.
•
Connect a piece of tubing to the squeeze bulb with bleeding screw; connect the opposite end to the square
fitting connector.
fitting connector.
•
Connect another piece of tubing to the square fitting connector; connect the opposite end to the calibrated
pressure meter.
pressure meter.
•
Connect another piece of tubing to the square fitting connector; connect the opposite end to the test volume.
•
Connect the last piece of tubing to the square fitting connector; connect the opposite end to the 'Y' fitting on
the two-tube blood pressure connector.
the two-tube blood pressure connector.
OpenTopic | ProBP 3400 calibration and verification setup | 9
Advertising
Popular Brands
Popular manuals
Advertising
Calibrate and verify the device
1. Click within the device row, then click Perform All to activate the test sequence for the module.
2. Enter information about any replacement parts used during service of the product. After entering the part
information, click Begin to start the tests.
3. Follow the onscreen prompts in the right side pane of the test window to progress through the test sequence.
The left pane shows the order of the required tests and tracks the test sequence progress. As the Welch Allyn
Service Tool application completes each test, the pass/fail indicator changes to green for pass or red for fail, and
the application proceeds to the next test.
Service Tool application completes each test, the pass/fail indicator changes to green for pass or red for fail, and
the application proceeds to the next test.
Welch Allyn Probp 3400 User Manual Pdf
4. To terminate the active test sequence at any time, click Abort. To terminate all tests, click Abort All.
Welch Allyn Probp 3400 Manual
After the Welch Allyn Service Tool application completes all the tests, it displays the pass/fail test results. After
you have finished reviewing the test results, click Close.
you have finished reviewing the test results, click Close.
On the Device information tab, the Welch Allyn Service Tool updates to the last successful date that calibration
was performed. The only way to change the last calibrate date of service is to perform all tests and all tests pass.
Any test failure will not change the calibration date. To see the updated calibration date, close the device tab and
reopen.
was performed. The only way to change the last calibrate date of service is to perform all tests and all tests pass.
Any test failure will not change the calibration date. To see the updated calibration date, close the device tab and
reopen.